BC Series Medical Area Gas Alarm BC-HG-IG-03-OVA
    Publish time 2018-09-29 22:08    
BC Series Medical Area Gas Alarm BC-HG-IG-03-OVA
Alarm System is a highly modular system utilizing the most advanced technologies in electronics and software. The modular design can implement limitless configurations to meet different monitoring requirements.Future expansion can be achieved through additional panels,electronics board and sensors.

BC Series Area Alarm is designed to accept input signals of pipeline gas pressure,have built-in RS485 communication port allows each BC Series alarm to be networked for remote monitoring.The physical data will be processed and displayed on site by the area alarms.

Product Features:


>Modular system configuration
>1 to 7 gas kinds available
>Pressure units are customizable (Psi, kPa, Bar, MPa, inHg, and mmHg )
>High/low alarm/warning limits are customizable
>Built-in RS-485 communication port for networking
>Universal sensor with fast connector for easy installation
>Touch mute button can be artificially muted
>Compact size with large four-digit LED numerical displays
>Dual color LEDs for system statuses
>All parameters can be field adjustable
>Labels can be customized upon request



Advantages Compared With Traditional Alarm:



Our Digital Area Alarm
Traditional Digital Alarm
The digital signal can accurately display pressure values
Only provide ultra low voltage digital signals
Electronic structure and difficult to damage 
Mechanical structure, easily damaged due to severe vibration 
Lower Environmental requirements
Vulnerable to changes in temperature, humidity, oxidation
High accuracy, accuracy class 1
Usually 1.6-level precision
Stable, long time accuracy
Prone to errors lead to false alarms
Positive and negative pressure general purpose
Connection errors may cause damage to the header
Easy wiring, sensor box set , can use the hub data transfer
A single header to wired cable connected to the alarm box

Connection Drawing:


Electrical and Physical Specifications:


Front Panlel Injection Molded Plastic



Case Body
Metal Alloy



Physical Dimension
1gas:240*230*65mm
2gases:340*230*65
3gases:440*230*65
4gases:540*230*65

5gases:640*230*65mm
6gases:740*230*65
7gases:840*230*65

Wall Opening
1gas:225*210*60mm
2gases:325*210*60
3gases:425*210*60
4gases:525*210*60

5gases:625*210*60mm
6gases:725*210*60
7gases:825*210*60

Power 
Input: 100~240 VAC



Analog Input
Input Type: (1) Single-ended, voltage




Working Range: 9VDC±5%



Relay Output
Gases: 1 output per 1 input gas




Ordering Information:


BC AA 03 OVA
Series Area Alarm Number of Gases Gas Type


(1-7 Normally) O: Oxygen



V: Vacuum



A: Medical Air



I: Instrument Air



N: Nitrogen



2: Nitrous Oxide



C: Carbon Dioxide


Example:

BC-AA-03-OVA indicates an area alarm for oxygen, vacuum, and medical air.